Stanhope Setaflash Series 8 Closed Cup Flash Point Tester – High Temperature - 82000-2
Setaflash Series 8 Closed Cup Flash Point Tester for High Temperature - 82000-2
The Setaflash Series 8 Closed Cup Flash Point Tester for high temperature (82000-2) delivers quick and dependable flash point measurements across a broad temperature range from ambient up to 300 °C, without the need for external cooling.
Using the Series 8, tests are completed in under 2 minutes with only a 2 or 4 ml sample. Its user-friendly interface, featuring a color digital display and touchscreen icons, simplifies the testing process. For convenient record-keeping, the instrument saves test results, which can be transferred to a laboratory information management system (LIMS) via a USB port.
Engineered for precision and accuracy, the device includes an internal barometer to measure barometric pressure and automatically applies corrections to the results. It also offers a ramp mode for testing unknown samples.
Key Advantages for You
The Setaflash Series 8 Closed Cup Flash Point Tester for high temperature (82000-2) delivers has a wide variety of advantages that you can benefit from:
The Setaflash Series 8 Closed Cup Flash Point Tester for high temperature (82000-2) is applicable in a wide range of industries including, but not excluded to:
Methods of the Setaflash Series 8 Closed Cup Flash Point Tester for high temperature (82000-2)
The Setaflash Series 8 Closed Cup Flash Point Tester (high temperature) is fulfilling various methods, which are listed below.
Methods
ASTM D3278
Standard Test Methods for Flash Point of Liquids by Small Scale Closed-Cup Apparatus - paints, enamels, lacquers, varnishes, and related products
ISO 3680 (Obs)
Methods of test for paints. Tests on liquid paints (excluding chemical tests). Test for flash/no flash (rapid equilibrium method)
ASTM E502
ASTM Standards for the Determination of Flash Point of Chemicals by Closed Cup Methods
ISO 3679
Determination of flash point. Rapid equilibrium closed cup method
ASTM D3828
Standard Test Methods for Flash Point by Small Scale Closed Tester
IP 523
Determination of flash point. Rapid equilibrium closed cup method
IP 524
Methods of test for paints. Tests on liquid paints (excluding chemical tests). Test for flash/no flash (rapid equilibrium method)
ASTM D7236
Flash Point by Small Scale Closed Cup Ramp Method
IP 534
Flash Point by Small Scale Closed Cup Ramp Method
ISO 9038
Determination of sustained combustibility of liquids
ASTM D1655
Standard Specification for Aviation Turbine Fuels
IP 602
Determination of sustained combustibility of liquids
CLP Regulations EC No 1272/2008
European Regulation (EC) No 1272/2008 on classification, labelling and packaging of substances and mixtures
BS EN ISO 3679 & BS 2000-523
Determination of flash no-flash and flash point. Rapid equilibrium closed cup method.
ASTM D8174
Standard Test Method for Finite Flash Point Determination of Liquid Wastes by Small-Scale Closed Cup Tester
SW-846 1020C
Standard Test Methods for Flash Point by Setaflash (Small Scale) Closed-Cup Apparatus
Quick Results and Simple Calibration & Verification
Results are determined in just one or two minutes, making it the fastest method for checking flash points. This efficiency maximizes sample throughput.
Furthermore, the software allows you to calibrate both temperature and barometric pressure, ensuring accuracy. Verification is further enhanced with Seta’s Small Scale Certified Flash Point material.
The statistical quality control (SQC) feature enables you to evaluate previous test results by calculating the mean, standard deviation, standard error, and repeatability. It also plots results according to ASTM D6299 standards. You can choose to analyze either all stored results or a selected subset of results.
User-friendly interface with color LCD display and an intuitive menu with touchscreen navigation. Moreover, there is a 1 GB internal memory.
This instrument also has a minimal sample requirement of 2 ml, which means lower costs and reduced waste per test for you. At the same time, it enhances your safety.
In the quest for the flash point, the Series 8 offers features like ramp mode for testing unknown samples, with a controlled increase of 2 °C per minute. The ActiveCool model integrates Peltier cell technology, ensuring swift and accurate heating and cooling without the need for an external water bath.
The Setaflash Series 8 Closed Cup Flash Point Tester for high temperature (82000-2) can be operated in 4 easy steps. Below you see the operator sequence.
What is flash point?
A test sample is placed in a heated cup equipped with a lid. An ignition source is introduced through an opening in the lid into the vapor space above the heated sample. Observations are made to determine if the vapor ignites (flashes) at the recorded temperature of the sample.
The determination of flash point is outlined in test methods established by standardization organizations such as the Energy Institute in the UK, ASTM in the USA, CEN in Europe, and ISO internationally.
Moreover, atmospheric pressure can influence the result. Low barometric pressure causes flammable vapors to form at a lower temperature, while high atmospheric pressure delays the formation of flammable vapors until a higher temperature is achieved.
Why should you test for flash point?
Flash point testing has been a well-established practice for over a century, helping to evaluate and classify the safety hazards of materials for classification, storage, and transportation. This testing is utilized across various industries, including petrochemical, waste management, and cosmetics.
The primary purpose of measuring flash point is to ensure that appropriate and specific precautions are taken during the manufacturing, storage, transportation, or usage of liquids.
Flammability is categorized into three main classes:
Extremely flammable: Flash point below 0°C
Highly flammable: Flash point below 23°C
Flammable: Flash point below 60°C
Changes in flash point can indicate the presence of hazardous volatile materials or the adulteration of a product. For instance, pure ethylene glycol has a flash point of 111°C, but with just 2% acetaldehyde, the flash point drops to 29°C. Similarly, adding petroleum spirit to engine oil will lower its flash point, indicating possible adulteration.
Accurate and reliable flash point measurements require standardized testing conditions, as over 20 factors can influence the result. These factors include the dimensions of the test cup and lid, the temperature and position of the ignitor, the dipping sequence and frequency, test duration, heating rate, atmospheric pressure, and more.
What is the difference between Close Cup and Open Cup?
Flash point is determined using either an open cup or closed cup flash point tester. Open cup tests are specified in some regulations and are designed to simulate conditions in open environments. In contrast, closed cup tests more accurately reflect situations where space is confined.
Let us explain the difference a bit more in depth. See below.
Closed cup tests are designed to simulate the scenario of a liquid spill in an enclosed environment. If the liquid is at or above its flash point, exposure to an ignition source could lead to a fire or explosion. In these tests, the sample is placed in a sealed (closed) test cup, and an ignition source is introduced to determine the temperature at which the sample ignites, referred to as its flash point.
Open cup tests are designed to simulate the potential ignition of a liquid spill in open conditions, such as a pool of liquid on the floor. These tests are used for fire point, combustibility, and sustained burning evaluations. The fire point is defined as the lowest temperature at which the liquid's vapors ignite and continue to burn after the ignition source is removed. Combustibility and sustained burning tests are typically conducted at a fixed temperature to assess the continuous burning of the sample.
An open cup tester generally yields a higher flash point compared to a closed cup tester because vapors can freely escape into the atmosphere. Closed cup tests are often preferred due to their greater precision.
The term non-equilibrium indicates that the vapor is not in equilibrium with the liquid. Non-equilibrium flash point tests involve heating the liquid at a steady rate while applying the ignition source at regular intervals.
Advantages:
Well-suited to automation
Universally available instrumentation
Standardized for a wide range of products.
Disadvantages:
Valuable components may escape each time the flame is dipped, leading to artificially high flash points or reduced precision
Significant temperature variation between the liquid and vapor in the cup
Requires a large sample size (50 to 80 ml per test).
Equilibrium
Setaflash
Traditional equilibrium flash point tests use any cup in a water bath to ensure that the liquid and vapor of the sample are in temperature equilibrium. This is achieved through a complex procedure and a very slow heating rate.
The Setaflash small-scale method simplifies the process by using a 2 ml or 4 ml sample, achieving rapid equilibrium between the vapor and liquid in a cup, and providing a reliable flash point result in just one or two minutes.
Advantages:
Improved precision compared to non-equilibrium methods
The Setaflash instrument offers low-cost, fast measurement for a flash/no flash result
Widely used by many industries.
Disadvantages:
Traditional testers can involve a laborious process, especially when larger sample sizes are needed
This type of instrumentation is not universally available.
Which Flash Point Testing Method should You be Using?
The suitable method for testing your sample is typically specified in a product specification or regulation, and in some cases, a referee test method may be indicated, which should be the preferred option. If multiple test methods are provided, the choice will depend on factors such as sample size requirements, testing speed, or precision.
When testing specifically for contaminants, some test methods and procedures are more suitable than others. Generally, an equilibrium test method is recommended for samples that might contain traces of volatile contaminants.
You can see the different Testing Methods below.
Seta Flash Small Scale
These methodologies encompass petroleum products and biodiesel liquid fuels, ranging from -30 to 300 °C, utilizing a compact closed cup tester.
These procedures can ascertain whether a product exhibits flash or not at a designated temperature (flash/no flash method A), or determine the flash point of a sample (method B). When coupled with an electronic thermal flash detector, these methodologies are also applicable for flash point assessments of biodiesel like FAME.
Small-scale testers are additionally employed in compliance with shipping and safety regulations such as CLP, aiding in the definition and classification of flammable and combustible materials.
These testers meet the following methods and regulations:
These testing procedures encompass the determination of the flash point of petroleum products within the temperature range of 40 to 360 °C, utilizing either a manual or automated PMCC apparatus. Additionally, they are employed to ascertain the flash point of biodiesel within the temperature range of 60 to 190 °C, utilizing automated PMCC apparatus.
These testers meet the following methods and regulations:
ASTM D93; IP 34; ISO 2719
Tag
These testing methodologies encompass the determination of the flash point using both manual and automated closed testers for liquids with a viscosity lower than 5.5 mm²/s (cSt) at 40 °C, or below 9.5 mm²/s (cSt) at 25 °C, and with a flash point below 93 °C.
These testers meet the following methods and regulations:
ASTM D56; ASTM D3934; ASTM D3941; ASTM E502; IP 491; IP 492; ISO 1516; ISO 1523
Cleveland
These testing procedures outline the process of determining the flash point and fire point of petroleum products using either a manual Cleveland open cup apparatus or an automated Cleveland open cup apparatus. They are applicable to all petroleum products with flash points above 79 °C and below 400 °C, excluding fuel oils.
These testers meet the following methods and regulations:
Why does Flash Point Autosamplers not Comply with Flash Point Test Methods?
Flash point autosamplers or carousels are not integrated into any flash point methods. Consequently, flash points obtained through an autosampler should not be relied upon for product classification concerning transportation, safety, health regulations, or specification purposes. Incorrect product classification can lead to serious repercussions.
Sample cups utilized in autosamplers fail to comply with flash point test method requisites due to the following reasons:
Samples contained within a carousel cup with a lid do not satisfy the criteria for a sealed container or the minimum sample volume as prescribed by the test method.
Test methods specify that sample volumes within containers that are less than 50% full can impact the accuracy of the test outcome.
Test methods caution that the loss of volatile materials can lead to the acquisition of inaccurately high flash points.
Furthermore, precision and bias have not been established for Autosamplers. Autosamplers have not undergone any flash point Interlaboratory Study Program (ILS). An ILS study enables the generation of precision and repeatability statements for a test method. Thus, any precision mentioned in flash point test methods does not pertain to autosamplers.
Data Sheet - Setaflash Series 8 Closed Cup Flash Point Tester for high temperature - 82000-2 - Stanhope
English | pdf
Article - Flash Point Testing Guide - Setaflash Series 8 Closed Cup Flash Point Tester for high temperature - 82000-2 - Stanhope
English | pdf
Article - Flash Point Autosampler vs. Flash Point Testing Explained - Setaflash Series 8 Closed Cup Flash Point Tester for high temperature - 82000-2 - Stanhope
English | pdf
Didn't you find what you were looking for?
If you are looking for some specific material, please write to us. Then we will try to find it for you, and send it as soon as possible.