Technical Specifications | |||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Medium | Water-based |
Sensor Data | |||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Measuring range | continuously adjustable from 0 to 20 cm/s to 0 to 200 cm/s |
||||||||||||||||||
Adjustibility | stepless from 10 % - 100 % by means of zero point and measuring range potentiometer |
||||||||||||||||||
Medium temperature | max. +135 °C |
||||||||||||||||||
Pressure | max. 30 bar |
||||||||||||||||||
Response time | 2 -10 sec. depends on measuring conditions |
||||||||||||||||||
Linearity deviation | < 5% most favourable straight line related to the final value |
||||||||||||||||||
Repeatability | < 2 % |
||||||||||||||||||
Temperature drift | < 0.3% pro Kelvin |
Mechanical data | |||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Protection class | IP 65 |
||||||||||||||||||
Material housing / electronics | Makrolon® |
||||||||||||||||||
Material housing / sensor | stainless steel AISI 303 |
||||||||||||||||||
Material of sensor head | Stainless steel AISI 303 (electro-polished) |
||||||||||||||||||
Material TC - closure | stainless steel AISI 316L (electro-polished) |
||||||||||||||||||
Sensor cable | 2 m encapsulated, shielded silicone cable 3 x 0,5 mm² |
||||||||||||||||||
Electrical connection | terminal block |
Electrical data | |||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
Operating voltage | 24 VDC ±10% |
||||||||||||||||||
Power consumption | approx. 100 - 200 mA (at max. flow) |
||||||||||||||||||
Ready for operation | approx. 10 sec. after power connection |
||||||||||||||||||
Output 4015.30 | 4 - 20 mA |
||||||||||||||||||
Load 4015.30 | max. 600 Ω |
||||||||||||||||||
Output 4015.31 | 0.1 - 10 V |
||||||||||||||||||
Load 4015.31 | ≥ 100 k-Ohm |
||||||||||||||||||
Protection circuit | reverse polarity & short circuit / overload protected up to max. 30 VDC only |
||||||||||||||||||
Output display | green LED |