Logo
Operation
Test methods

ASTM D92; D8254; IP 36; ISO 2592; DIN 51 376; NF T60-118; JIS K 2265-4; AASHTO T48

Test modes

Expected flash point, fire point and search mode

Ignition system

High purity platinum electric hot-wire or gas flame; automated gas flame relight; automatic electric ignitor failure detection

Flash detection

Ionisation ring

Fire detection

Ionisation ring and optical fire sensor

Cooling

Fast forced air (integral fan, adjustable)

Heating rate

Adjustable fast heating mode (up to 17 °C/min) and standard rate 5.50 °C/min

Calibration

Calibration and verification dates/data stored, easy retrieval and password protected

Application range

Ambient to 400 °C

Draught shield

Included, draughts protected on 3 sides, front and side move down for easy access

Measurement
Sample temperature probe

Class A PT 100 stainless steel probe, 0.1 °C resolution, -50 °C to +450 °C

Resistance calibration with unlimited data points (more than 5 is recommended)

Units of temperature

°C or °F (user selectable)

Barometric pressure

Automatic flash point correction with built-in pressure sensor, kPa (incudes alternative units such as hPa, mbar, mmHg)

Data Management
Information

Real-time display on screen of test progress and results, results securely stored in a SQL database

Internal memory

More than 7 Gb of storage

Statistical quality control (SQC)

Built in SQC module according to ASTM D6299 to analyse SQC samples

Safety
Fire extinguisher

Snuffer plate is automatically deployed at the end of test to extinguish fires in sample cup

Optical sensor deploys inert gas* around cup and cup area if persistent fire is detected
*requires a supply of inert gas (N2, Argon etc.) at 140 kPa to 160 kPa

Manual fire extinguisher deployment switch is available as an option, and can be mounted at a discrete distance, outside the fume cupboard to enable a user to deploy the fire extinguisher manually if required

Over temperature cutout safety device

>400 °C

Gas Supply (optional)
Gas type

Butane, propane or natural gas source

Gas supply

30mbar (3kPa)

Power Requirements
Power

100 V to 240 V, 47 Hz to 63 Hz, autosensing.

Maximum 600 W, during typical test <100 W

Interface Specifications
Display

Large 18 cm (7 inch) LCD colour touchscreen, 1024 x 768 pixels

Data input/output

Connection to LIMS via Ethernet RJ45, USB Type A (Qty 2), test results can be exported to memory stick in various formats (PDF or CSV) or transferred via QR code

Printer options

USB, Ethernet or RS232C

Language

User selectable - English (default)

Environmental conditions
Operating temperature

5 to 35 °C (50 to 104 °F)

Relative humidity

Up to 80% at 35 °C (not condensing)

Altitude

2000m maximum

Physical
Dimensions (HxWxD)

240 x 385 x 450 mm

Weight

15.5 kg

To top

Search