Logo
Specifications
Measurement
  • Oil types: As Per MEPC 108(49) + MEPC 240(65) requirements
  • Clean water calibration: Automatic
  • Oil measurement range: 0 - 1000 ppm all types
  • Resolution: 1 ppm
  • Accuracy oil + solids: As Per MEPC 108(49) requirements
Data Storage and Retrieval
  • Data retrieval:  via LCD display or download to PC using Hyperterminal
System and Supply
  • Supply voltage: 115 / 230V ac, 50 - 60Hz (Switchable)
  • Zener Barrier/ Computer Module: 115 / 230V ac, 50 - 60Hz (Switchable)
  • Motor: 380-440V ac, 50-60Hz, 3 phase, 250W
  • Supply voltage Consumption: > 50 VA Single Phase
  • Approvals:  mEPC 108 (49) - DNV GL, GL + USCG, ABS, CCS, NKK, BV, and Russian Register. As well as, MEPC 240(65) - DNV GL
To top