Logo
Technical Specifications
Explosion-proof Grade
  • ATEX & IECEX: Ex ia IIC T4 Ga (Ta = -20 °C to +50 °C)
  • CSA:
    Class I, Division 1, Group A,B,C and D, T4
    Class I, Zone 0, AEx ia IIC T4 Ga
    Ta = -20°Cto + 50°C Ex ia IIC T4 Ga
  • UKCA-EX
CE Certification

TUV IEC61326, IEC61010

Input Channels
  • Top: 2 channels RTD measurement, 1 channel electric signal measurement, φ4mm banana jacks
  • Right side: 2 channels for external digital pressure module, LEMO style connection
  • Bottom: embedded digital pressure module (model ADT158Ex), field switchable.
  • Internal: 1 embedded atmospheric pressure sensor
Loop power
  • 20V ± 10%
  • max output impedance: 320Ω
  • max load current: 25mA
Measurement Rate
  • mV, V, mA & frequency and RTD: 3 times/ sec
  • Pressure module: 1~10 times/sec selectable (3 as default)
  • Barometer: 1 time/sec
Range:

RTD:

  • 0...400 Ω

Voltage:

  • ±300 mV
  • ±30 V

Current:

  • ±30 mA

Frequency:

  • 0,01...50000 Hz (auto range)

Pressure:

  • From -1...4200 bar
Accuracy

RTD:

  • 0,01 %RDG + 0,005 %FS

Voltage:

  • 0,015 %RDG + 0,005 %FS
  • 0,015 %RDG + 0,005 %FS

Current:

  • 0,015 %RDG + 0,005 %FS

Frequency:

  • 0,005 %RDG + 2 sidste cifre

Pressure

  • 0,02 ... 0,05 %
Resolution

RTD:

  • 1 mΩ

Voltage:

  • 1 uV
  • 0,1 mV

Current:

  • 0,1 uA

Frequency:

  • 6-digit auto-resolution
Data Storage

Logging interval: from 0.1~9999 seconds, log up to 10 million readings (single channel)

To top